Click on image to view larger version.

Fig. 7. Positioning the shunting electrodes farther away from the side of the fish
reduces the current shunted during an electric organ discharge (EOD) but
leaves the distribution of responsiveness across phases constant. (A)
Recording of EOD current shunted during prolonged switch closure when the
shunt electrodes were placed at three lateral displacements, successively
farther away from the fish cage (0, 1 or 2 cm). Each point of the waveforms
predicts the magnitude of EOD current that is shunted when the electronic
switch is closed during an interval that centers at this point. An example of
the headtail EODs, recorded simultaneously, is shown below as a phase
reference. (B) Response probabilities obtained as the switch was closed for
100 µs either during one of the extrema V2,
V3 or V4, or 4 ms after onset of the
EOD (Control). For each of the three distances, 100 responses were recorded at
each of the four stimulus conditions (total of 1200 tests on fish 6). The
baseline response probability (due to spontaneous rate fluctuations) is given
by the dotted horizontal line (pooled from 100 recordings without switch
closure at each distance). There was no significant difference between
baseline and controls. At all distances, responsiveness in
V4 is significantly higher than in V3
(P<0.001) and higher in V3 than in
V2 (P<0.05). All response levels are
significantly above baseline except for those at distances 1 and 2 cm in phase
V2. Decrease in responsiveness with distance is
significant in V4 (0-1 cm: P<0.01) and
V3 (0-2 cm: P<0.05).