Journal of Experimental Biology 61,667-675 (1974)
Published by Company of Biologists 1974


A Method for Determining Total Carbon Dioxide in Nanolitre Volumes of Liquid

COLIN LITTLE 1

1 Department of Zoology, University of Bristol, Bristol BS8 1UG

1. A microdiffusion method is described by which total carbon dioxide can be determined in 2 nl samples of liquid. The standard deviation for 0-30 mM/l HCO3- is approximately 1 mM/l.

2. The volume of sample can be reduced to o·8 nl, but the S.D. then rises to 1·6 mM/l. With volumes of 0·4 nl, the S.D. is 2·7 mM/l. It is not possible, without modification, to use volumes smaller than about 0·4 nl.

3. In samples where the HCO3- concentration is low, accuracy can be increased by using larger volumes: 10 mM/l HCO3- can be determined in 6 nl samples with a standard deviation of 0·5 mM/l.

Submitted on June 8, 1974







© The Company of Biologists Ltd 1974