spacer gif spacer gif spacer gif spacer gif spacer gif
 QUICK SEARCH:   [advanced]


spacer gif
     Home     Help     Feedback     Subscriptions     Archive     Search     Table of Contents    

This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by RAMSAY, J. A.
Right arrow Articles by FALLOON, S. W. H. W.
Right arrow Search for Related Content
PubMed
Right arrow Articles by RAMSAY, J. A.
Right arrow Articles by FALLOON, S. W. H. W.
Journal of Experimental Biology 30,1-17 (1953)
Published by Company of Biologists 1953


Simultaneous Determination of Sodium and Potassium in Small Volumes of Fluid by Flame Photometry

J. A. RAMSAY 1, R. H. J. BROWN 1, and S. W. H. W. FALLOON 1

1 The Zoological and Cavendish Laboratories, University of Cambridge

1. A method of flame photometry is described, by which the amount of sodium and potassium in biological fluids can be determined simultaneously, using samples of the order of 10-3 cu.mm.

2. When the method is tested with quantities near the lower limits of detection (of the order 4 x 10-8 mg. sodium, 6 x 10-8 mg. potassium) the reproducibility, measured as standard deviation, is ±7 x 10-9 mg. sodium and ± 17 x 10-9 mg. potassium.

3. When larger quantities (up to certain limits) are used, the standard deviation is approximately ± 3 % of the quantity in the sample, both for sodium and for potassium.

4. As is usual in flame photometry serious errors can be caused by the presence of other substances in the sample. In the case studied for purposes of illustration it is shown that these interference errors can be reduced to the order of ± 6 % by the addition of excess ammonium phosphate.

Submitted on May 29, 1952




This article has been cited by other articles:


Home page
ScienceHome page
G. G. Vurek and R. L. Bowman
Helium-Glow Photometer for Picomole Analysis of Alkali Metals
Science, July 23, 1965; 149(3682): 448 - 450.