Fig. 3. Control trace demonstrating successful identification of a thermal conversion factor for switching between N2 and He using a single mass-flow controller. The stream flowing through the CO2 analyzer was composed of 5 ml min–1 of a span gas (505 p.p.m. CO2 in N2), 21 ml min–1 O2 and 74 ml min–1 of either He or N2 carrier. Using the identified conversion factor (1.37), we were able to switch between carrier gases without affecting the reading from the CO2 analyzer (~25 p.p.m. CO2, which is 5% of 505 p.p.m.).