Fig. 3. Analysis of fluorescent fluid samples collected in optically flat hollow
rectangle glass capillaries. (A) XY image of a sample
of 50 µmol l1 Texas Red collected under paraffin oil in a
hollow rectangle glass capillary. The width of the image is
250 µm.
(B) XZ image of a sample of 125 µmol
l1 Texas Red. The solid line was positioned at the apparent
midline of the bright band of fluorescence, corresponding to the
Z-axis midpoint. For comparison, the upper and lower dotted lines
indicate positions 10 µm from the apparent Z-axis midpoint. The
dotted lines are clearly well away from the midline of the bright band in the
XZ image and are shown to indicate that the
Z-axis midpoint can be reliably judged by eye. (C) Fluorescence
intensity of XY images collected for the same sample
at the Z-positions indicated. The curve shows that fluorescent
intensity is nearly constant when the optical slice is positioned within
5 µm of the apparent Z-axis midpoint. (D) Representative
calibration curve relating fluorescence intensity to the concentration of
Texas Red in fluid samples collected in optically flat glass capillaries.
Texas Red concentrations are plotted on the Y-axis so that the
equation of the curve fit to the data by non-linear regression analysis allows
calculation of the dye concentration of experimental samples from the measured
fluorescence intensities. Measured fluorescence intensities were divided by
100 for convenience of the curve-fitting calculations. For the curve shown,
CLSM detector gain was adjusted so that a near maximal signal (
3900 on a
scale of 04095) was obtained from the highest concentration (250
µmol l1) of Texas Red in this series of four calibration
samples.